Open the mobile version, with barcode scanning
Class II Terminated

DIANEAL Low Calcium (2.5 mEq/L) Peritoneal Dialysis Solution with 1.5% Dextrose, 5000 mL container, Rx only, Baxter Healthcare Corporation, Deerfield, IL 60015, Product L5B4826, NDC 0941-0409-07

Why: Defective Container: There is a potential for frangible components to be broken, resulting in a leak at the port when the closure is removed.

Company
Baxter Healthcare Corp.
Deerfield, IL, United States
Reported
Aug 7, 2013
Recall started
Jul 24, 2013
Ended
Jun 24, 2014
Where sold
Nationwide and Puerto Rico
Amount
39,486 containers
Full product description
DIANEAL Low Calcium (2.5 mEq/L) Peritoneal Dialysis Solution with 1.5% Dextrose, 5000 mL container, Rx only, Baxter Healthcare Corporation, Deerfield, IL 60015, Product L5B4826, NDC 0941-0409-07
Lots / codes affected
Product Code: L5B4826, Lot #: C886820, Exp 11/14

Class I: could cause serious harm or death. Class II: could cause temporary or treatable problems. Class III: unlikely to cause harm.