Class III
Terminated
Varithena (polidocanol injectable foam) Administration Pack, Contains: 3 silicone-free syringes, 2 compression pads, 1 Varithena Transfer Unit, 1 manometer tubing, Rx Only, Distributed by Biocompatibl…
Why: Defective Delivery System: incorrect silicone oil-free NormJect 10 mL Luer Lock Solo syringes packaged in the pack, instead of the required silicone oil-free NormJect 10 mL Luer Solo syringes (luer slip connection).
- Company
- Biocompatibles UK, Ltd.
Farnham, United Kingdom - Reported
- Dec 18, 2024
- Recall started
- Aug 15, 2024
- Ended
- Feb 21, 2025
- Where sold
- Nationwide in the USA.
- Amount
- 432 administration packs
- Full product description
- Varithena (polidocanol injectable foam) Administration Pack, Contains: 3 silicone-free syringes, 2 compression pads, 1 Varithena Transfer Unit, 1 manometer tubing, Rx Only, Distributed by Biocompatibles Inc., a BTG International Group company. CN01114.3
- Lots / codes affected
- Lot # 34067418, Exp. March 2026, 34067419, Exp. March 2026
Class I: could cause serious harm or death. Class II: could cause temporary or treatable problems. Class III: unlikely to cause harm.